Table 2-1R: US-EPA Method NPDES Standard Method

US-EPA Method NPDES (A22, 23)

Closed Vessel Microwave Digestion of Wastewater Samples for Metals Determination

Type

Matrix

Analytes

Detection

Control

Microwave

Acid Leach

Domestic and Industrial Wastewater

Al, As, Ba, Cd, Cr, Cu, Fe, Mn, Ni, Pb, Sb, Se, Zn

FAAS

ICP-AES

DCP-AES

Microwave

Unit Specific

Power Settings

Condensed Procedure

50.0 mL sample, 3 mL HNO3, and 2 mL HCl

For 1 - 6 samples:

For MW units with 364 - 420 W use 75% power for 30 min to a maximum temperature of 164 ± 4°C

For MW units with > 364 - 420 W at 75%, adjust power setting to deliver 364 - 420 W for 30 min.

For 7 - 12 samples:

For MW units with 575 - 635 W use 100% power for 50 min to a maximum temperature of 165 ± 5°C.

For MW units with 635 - 700 W use 100% power for 30 min to a maximum temperature of 165 ± 5°C.

For MW units with > 700 W adjust power settings to reproduce temperature versus time profiles illustrated in the method.