Table 2-1P: US-EPA Method 3052 Standard Method

US-EPA Method 3052 (A20)

Microwave Assisted Acid Digestion of Siliceous and Organically Based Matrices

Type

Matrix

Analytes

Detection

Control

Microwave

Total Digestion

Fly Ash

Oil

Sediment

Sludge

Soil

Al, Ag, As, B, Ba,

Be, Ca, Cd, Co,

Cu, Cr, Fe, Hg, K,

Mg, Mn, Mo, Na,

Ni, Pb, Sb, Se, Sr,

Tl, V, Zn

FAAS

ET-AAS

ICP-AES

ICP-MS

Temperature

Feedback

Control

Condensed Procedure

Up to 0.5 g sample and 9 ± 0.1 mL HNO3 and 3 ± 0.1 mL HF (variable up to ± 2 mL) and alternatively 2 ± 2 mL HCl (can not be analyzed by ET-AA)

Performance Criteria:

Heat to 180 ±5°C in less than 5.5 min.

Maintain 180 ±5°C for remainder of 15 min